Cervical consistency has traditionally been assessed by digital examination, which remains the standard method despite its limited performance. Elastography offers a more objective way to evaluate cervical biomechanics, and its results can reach the clinician as a color map or a value in meters per second or kilopascals. Interpreting a result correctly starts with understanding what the test measured and how.
What Does Elastography Measure
Elastography estimates tissue stiffness. Depending on the stimulus that deforms the tissue, conventional elastography includes two ultrasound-based techniques:
Strain elastography (SE) uses probe compression or physiological motion to deform the tissue. Under an equal amount of stress, stiffer regions deform less than the surrounding soft tissue. The output is a color map coding different magnitudes of strain that can be superimposed on the conventional ultrasound image. SE is a qualitative technique displaying only relative stiffness with no absolute elastic modulus.
Shear wave elastography (SWE) estimates stiffness from the propagation speed of a shear wave. When this wave passes through the tissue, the velocity of the wave varies with tissue stiffness. Contrary to SE, SWE is a quantitative technique, providing a numerical result expressed in meters per second or converted, under certain assumptions, into kilopascals.
The Elastography Scale: Stiffness Index Values and Color Mapping
The color map is the most familiar output. Depending on the manufacturer, the color map shows a scale from red to blue indicating soft to stiff, but the reverse setting can also be used, so the labels of the displayed color bar must be checked before any interpretation. If a grayscale map is used, it should not be displayed over the grayscale B-mode image, but rather as a separate strain image.

Sagittal view of a cervix at gestational week 12 with a color box superimposed. The anatomic area of interest is marked with red. *Artifact in the anterior vaginal fornix. Image credit: Thomsen et al. 2024. CC BY 4.0.
Semiautomatic software such as E-Cervix converts the color distribution of the elastogram into quantitative parameters, including the hardness ratio (HR), the internal and external os strain (IOS, EOS), and an elasticity contrast index (ECI) that reflects tissue heterogeneity. These values are derived from the color map, not measured directly.
Shear wave elastography, by contrast, provides a quantitative estimate: it measures the propagation speed of the shear wave in meters per second, which can be converted into Young’s modulus in kilopascals under certain assumptions about the tissue. That number, however, has no reference tissue in the case of the cervix.
No clinical guidelines establish reference values for cervical stiffness measured by SWE, either during or outside of pregnancy. The values reported by individual studies cannot serve as a reference either. SWE values vary between machines, and in the cervix they also differ by region and with gestational age.
Interpreting Results: Threatened PTL and Induction Readiness
Cervical stiffness is clinically relevant because the cervix changes its composition before it changes its shape. It must remain firm throughout pregnancy and soften at term; preterm cervical ripening may cause preterm birth, while the absence of ripening at term is associated with an increased risk of unsuccessful labor induction. These biochemical changes precede morphological change, which cervical length alone cannot capture. In a preliminary study of 18 pregnant women, cervical stiffness decreased progressively across gestation but showed no significant correlation with cervical length, suggesting that the two measures carry different information (Massó et al., 2019).
In threatened preterm labor (TPTL), this softening is measurable. In a retrospective cohort of 95 singleton pregnancies presenting to obstetric triage between 23+0 and 33+6 weeks, women who went on to deliver preterm had a softer cervix on every E-Cervix parameter (Nazzaro et al., 2022).
Changes in cervical consistency can also help predict induction readiness. A 2016 systematic review and meta-analysis in the Journal of Perinatal Medicine, pooling four studies and 323 women, found that cervical elastography shows superior accuracy compared to the Bishop score. It could be used not only for a more reliable cervical stiffness assessment, but also for a better prediction of labor induction success (Londero et al., 2016).
These studies are examples of cervical stiffness carrying clinically relevant information, but none of the published values can yet be applied as a universal clinical decision threshold.
Common Pitfalls in Conventional Elastography Interpretation
Many errors in conventional elastography are introduced during acquisition and then carried silently into the final value. For strain elastography, scans of tissue simulators and pregnant women have identified several of them. Strain values decrease as the region of interest (ROI) moves further from the probe and deviates from the probe centerline. Soft tissue between the probe and the ROI makes the posterior lip appear falsely harder than the anterior lip (Thomsen et al., 2024).
ROI placement is a particularly weak point. The authors recommend placing it in the middle third of the anterior cervical lip, yet many published studies deviate from this, and selecting the anatomic area of interest proved especially problematic. Pre-compression adds a further source of error, as excessive probe pressure is the most likely cause of inconsistent, non-reproducible elastograms.
Shear wave elastography has its own sources of error. In a study of 154 pregnant women, depth from the probe did not significantly affect shear wave speed, but values differed between the internal os, the external os, and the endocervix, so ROI placement still matters (Hernandez-Andrade et al., 2014). In addition, converting wave speed into stiffness also relies on assuming a homogeneous, non-viscous, isotropic medium, which fits poorly with a cervical stroma formed of cross-linked collagen immersed in viscous proteoglycan. Commercial shear wave transducers also struggle to isolate a pure shear signal from longitudinal waves, which can produce multiple reflections in a small organ like the cervix.
CerviSense is built on a different measurement principle. Torsional wave elastography (TWE) does not rely on ultrasound imaging, and its waves propagate in a curled pattern, radially as well as in depth, a feature that makes the technique suitable for cervical tissue.
In our intravaginal probe, a rotational actuator and a piezoceramic receiver ring transmit and receive the waves, and the result is a numerical stiffness value. This value is combined with condition-specific clinical variables through AI models currently in clinical validation for threatened preterm labor, preterm labor screening, and induction of labor.
The intravaginal device is already CE-marked, and it is listed in the Voluson™ Solution Store by GE HealthCare. In a reproducibility study of 77 pregnant women across five Spanish hospitals, we reported an intraobserver ICC of 0.88 for the device. After adding a lateral microcamera and operator training, interobserver ICC reached 0.93 in 16 additional women (Molina et al., 2023).

We are also building evidence to set a validated reference for normal cervical stiffness. The CerviSense Norm study will kick off in the last quarter of 2026. Asymptomatic women with singleton pregnancies between 18 and 22 weeks will have their cervical stiffness measured as part of routine care. The goal is to build a reference curve for cervical biomechanics.
Reproducibility is the foundation on which clinical thresholds can be built. Establishing those thresholds for CerviSense and demonstrating its clinical utility is the next step.
Sources
Dietrich CF et al. Strain Elastography – How To Do It? Ultrasound Int Open. 2017 Sep;3(4):E137-E149. doi: 10.1055/s-0043-119412.
Hernandez-Andrade E, Aurioles-Garibay A, Garcia M, et al. Effect of depth on shear-wave elastography estimated in the internal and external cervical os during pregnancy. J Perinat Med. 2014;42(5):549–557. doi: 10.1515/jpm-2014-0073
Kim H, Hwang HS. Elastographic measurement of the cervix during pregnancy: Current status and future challenges. Obstet Gynecol Sci. 2017 Jan;60(1):1-7. doi: 10.5468/ogs.2017.60.1.1.
Londero AP, Schmitz R, Bertozzi S, Driul L, Fruscalzo A. Diagnostic accuracy of cervical elastography in predicting labor induction success: a systematic review and meta-analysis. J Perinat Med. 2016;44(2):167–178. doi: 10.1515/jpm-2015-0035
Massó P, Callejas A, Melchor J, Molina FS, Rus G. In vivo measurement of cervical elasticity on pregnant women by torsional wave technique: a preliminary study. Sensors. 2019;19(15):3249. doi: 10.3390/s19153249
Mlodawski J, Mlodawska M, Plusajska J, et al. Repeatability and reproducibility of quantitative cervical strain elastography (E-Cervix) in pregnancy. Sci Rep. 2021;11:23689. doi: 10.1038/s41598-021-02498-3
Molina FS, Pardo L, Muñoz MD, et al. Reproducibility and usability assessment of the novel Fine Birth device for threatened preterm labor diagnosis. Am J Obstet Gynecol MFM. 2023;5(7):100982. doi: 10.1016/j.ajogmf.2023.100982
Nazzaro G, Saccone G, Miranda M, Crocetto F, Zullo F, Locci M. Cervical elastography using E-cervix for prediction of preterm birth in singleton pregnancies with threatened preterm labor. J Matern Fetal Neonatal Med. 2022 Jan;35(2):330-335. doi: 10.1080/14767058.2020.1716721.
Thomsen CR, Jensen MSS, Bor P, Hinge M, Sandager P, Uldbjerg N. Recommendations for strain elastography of the uterine cervix. Arch Gynecol Obstet. 2024 Oct;310(4):2023-2033. doi: 10.1007/s00404-024-07693-x.